ORIGINAL ARTICLE
Year : 2013  |  Volume : 8  |  Issue : 3  |  Page : 159-166

Evaluation of the technical specifications of computerized tomography scanners in Jazan


1 Department of Radiology, Jazan University, College of Applied Medical Science, Jazan, KSA
2 College Medical Radiological Science, Sudan University of Science and Technology, Khartoum, Sudan
3 National College of Medical and Technical Studies, Khartoum, Sudan

Date of Web Publication16-May-2014

Correspondence Address:
Caroline Edward Ayad
Department of Radiology, 1College Medical Radiological Science, Sudan University of Science and Technology, Khartoum, Sudan

Login to access the Email id


DOI: 10.4103/1858-5000.132611

Rights and Permissions
  Abstract 

Introduction : good quality management for computerized Tomography (CT) scanners is essential to safe and efficient CT units, providing quality clinical images, maintaining patient and staff radiation doses as low as reasonably achievable. Aims : to evaluate the technical specifications of (CT) scanners in Jazan region in the period from 2011-2013. Materials and Methods : 13 CT scanners have been evaluated; 2 of them are in private sectors and the rest in public hospitals. The Technical specifications of CT scanners were assessed using template issued by ImPACT (Imaging Performance Assessment of CT scanners). Results : When comparing the 11 public scanners age with guidelines rules of European Coordination Committee of the Radiological and Electro medical Industries (ECCREI); it showed that scanners of Jazan region are within lifecycle guidelines, the total cumulative number of scanners since 1984 to 2013 are 15 scanners, 4 of them were replaced and the rest under use, multi detector CT scanners replaced most of the single detector scanners. for public CT scanners ; results show that all of the scanners are 3 rd generation, gantry bores are arranged between70-80cm, the x-ray tube inventory showed that there is no dual source CT scanner in the region and the anode storage heat capacity ranged from (3-8MHU) except Siemens 64slices and 20slices reached up to (30MHU). All of scanners in the region are built in solid state, image reconstruction time display per second is ranged from 1-40slice/seconds, advance clinical application software are available among the scanners. Jazan region CT scanners have a high capability and their technical specifications are in a rapid pace in developments that impacting on performance which depends on trade-off between image quality and patient dose.

Keywords: Computerized tomography scanner, Jazan, purchasing, quality management


How to cite this article:
Mohamed Ali ST, Hamad MM, Ayad CE, Abdalla EA, Ahmed AS. Evaluation of the technical specifications of computerized tomography scanners in Jazan. Sudan Med Monit 2013;8:159-66

How to cite this URL:
Mohamed Ali ST, Hamad MM, Ayad CE, Abdalla EA, Ahmed AS. Evaluation of the technical specifications of computerized tomography scanners in Jazan. Sudan Med Monit [serial online] 2013 [cited 2017 Apr 19];8:159-66. Available from: http://www.sudanmedicalmonitor.org/text.asp?2013/8/3/159/132611


  Introduction Top


The use of computerized tomography (CT) for medical diagnosis has increased over the past decades, resulting in increasing patient radiation dose from this imaging modality. [1],[2] The introduction of 64-slice CT scanners has further increased the patient throughput and the application for CT. There is a growing concern that radiation exposure from diagnostic imaging will increase the risk of cancer. [3]

The National Council on Radiation Protection in the USA has estimated that the contribution of medical radiation to the population collective dose has increased from 15% in 1980 to 53% in 2006, with CT accounting for the major share of this increase at approximately 1.5 mSv/capita/year. [4] Similar increases have been noted in the United Kingdom and this trend is common to most developed countries. The introduction of spiral CT scanners in 1989 and subsequently multislice CT detectors in 1998, represented major highlight in the development of CT scanners. [5]

Jazan region is a smallest region in south Saudi Arabia, population are 1,300,110 (Census 2010), 23% of the CT scanners are found in the king Fahad central hospital (KFCH); this is due to hospital size and rush of work as the hospital contains almost all of the medical specialty and the other hospital transfer most of the critical cases to (KFCH). The remaining 77% of the scanners are distributed between Sabia, Abu Arish, Jazan, Samtah Biesh, Eldarb, and Feyfa hospitals as well as 2 CT units divided between Alomies private hospital and Garash polyclinic. [6]

There are many publications in comparative technical specifications for CT scanners marketing, [7],[8],[9] which are useful for suppliers, buyers, vendors and operators of scanners, specially that publications done by the center for evidence - based purchasing (CEP) is the part of national health services purchasing and supply agency (UK), role of CEP is to help buyers in technical, operational and economic considerations. [9]

For continuous quality improvement of CT scanners, CT Service delivery should meet or exceed the needs and expectations of patients and staff. A good quality management for CT scanners is essential to a safe and efficiently run CT units, providing quality clinical images while maintaining patient and staff radiation doses as low as reasonably achievable. CT purchasing specifications is the first step in continuous quality improvement of CT scanners.

The aim of this study is to evaluate and to compare the technical specifications of CT scanners in Jazan region.


  Materials and Methods Top


This study carried out in Jazan region in kingdom Saudi Arabia, during the period from July 2012 to July 2013, it consists of 14 provinces.

Methods of assessment

In Jazan region there are 13 CTs in clinical use; 2 of them are in private sectors and the rest in public hospital including: GE Hispeed 4 slice, Siemens Somatom Definition 64 slice, 2 GE Light speed 16 slice, GE bright speed 16 slice, Siemens Somatom Emotion 6 slice, Toshiba Aquilion 16 slices, GE bright speed 8 slices, Siemens Somatom 20 slices, Siemens emotion 1 slices and neurologica ceretom 8 slices (portable scanner), all have been evaluated.

The technical characteristics of the CT scanners were assessed by using CT manuals and retaining to CT units in the region suring of the performance from the operators. Technical specification was assessed using template issued by Imaging Performance Assessment of CT scanners, by comparing of the scanners Gantry, X-ray tube, Generators, detectors design, couch, control, console, computer, storage media, workstations, clinical software and clinical application.

CT scanners age in the region compared against guidelines published by the rules of European Coordination Committee of the Radiological and Electro medical Industries (ECCREI) for the guidelines of CT scanners age published by the ECCREI According to the rules from ECCREI. [10]


  Results Top


Analysis CT scanners in the region show that multislice CT scanners comprise of 10 (77%) of the total; where single slice CT scanners were 3 (23%) of the total.


  Discussion Top


In Jazan region the 11 public scanners are distributed among 9 hospitals (19 public hospitals) 3 scanners are found in the central hospital (KFCH) as the central hospital almost include the whole specialty. The 13 scanners in clinical use in the region (until May 2013), that is approximately equal one CT per each hundred (100) thousands inhabitants (Population in region are 1,300,110 Census 2010), this proportion is high world-wide. [11] The majority of CT scanners (85% of the total) are in public hospital while the rest (15% of the total) are in private sectors [Table 2] and [Table 3].

This study presents the comparative technical specifications for public scanners only because of the un -cooperation of private sectors in the region. [Table 1] presented the age of Jazan's Public CT scanners inventories, relative to the rules from the European Coordination Committee of the Radiological and Electro medical Industries (ECCREI) May, 2013 The first CT in the region was installed in 1984 at KFCH; it was Siemens Somatom single slice, the cumulative number of public CT scanners installed (1984-2013) in the region are 15 scanners, 11 scanners of them are under use and 4 CT of the total are replaced, within this period X-ray tubes are changed for estimated 10 times for the total of the 15 scanners (replaced plus in use Scanners). [6]

The result of technical specification comparison for public CT scanners in the Jazan region show that all of the scanners are 3 rd generation as the 4 th generation are not not widely used by the manufacturers due to difficult to stabilise rotation and expensive detector. [5]

Slice class are arranged from 1 single slice (%), 4 (%), 6 (%), 8 (%), 16 (%) which are the most abundant, as well as 20 (%) and 64 (%), which is the advanced scanner recently in the region.
Table 1: The age of Jazan's public CT scanners inventories, relative to the rules from the ECCREI May, 2013

Click here to view
Table 2: Inventory of public CT scan in Jazan region

Click here to view
Table 3: Inventory of private CT scan in Jazan region

Click here to view


Corresponding to world-wide rapid pace of change in the provision of CT in recent years, in relation to detectors racing between CT manufacturers, Jazan region also show a rapid change in most of CT scanners.

The inventory showed that the gantry bores are arranged between 70 cm, 78 cm and 80 cm, there is availability in the future for radiation therapy planning for big bore CT for 80 cm gantry bore in GE scanners, there is some exception for Ceretom scanner which has 32 cm diameter as this portable CT is used only for head and neck scanning, the same Table also shows that the gantry tilt range for all of scanners are ±30, the max scan field of view for all scanners is 50 cm, the gantry rotation times ranged between 0.33 and 4 s both Siemens Somatom 64 slice and Siemens Somatom 20 slice are the most fastest among scanners in rotation time with range of (0.33-1.5 s) [Table 4].
Table 4: Technical specifi cation comparison (Gantry)

Click here to view


In the X-ray tube [Table 5], the inventory showed that there is no dual source CT scanner in the region. And the Anode storage heat capacity for the available scanners is ranged from 3 MHU to 8 MHU except Siemens somatom definition 64 slices and 20 slices in which the X-ray tube the anode storage heat capacity reached up to (30 MHU), The max anode cooling rate for scanners are ranged from 635 to 7300 in (KHU/min) except the portable scanner, which has 12 min maximum.
Table 5 Technical Specifi cations Comparison (X-Ray Tube)

Click here to view


Most of the tube changing took place in scanners more than 5 years age. Consumption of X-ray tube may give assign for overload of cases. [6] For the Guaranteed tubes life which are variable from scanner to other they depend on the contract; either annually or by number of scans/day/year/rotation [Table 5].

The X-ray generator of all scanners are located in the gantry and are using air as cooling method, Generator max output power rating is ranging between 42 and 100 kW and the kVp range available for all scanners is (80, 100, 120, 140 kv), in detector system [Table 6] all of our scanners in the region are build in solid state i.e., there are no gas detectors which are used mainly with the 4 th generation, all Siemens scanners used ultra-fast ceramic as detector material while GE scanners used highlight lumex, image reconstruction time display per seconds is ranged from 1 slice/s up to 40/s for Siemens 64 slices, Toshiba has the large number of detectors along the (z) axis and so large number of detection channels per row between the scanners, minimum slice thickness of scanners in the region are ranged from 0.5 to 1.5 mm.
Table 6 Technical Specifi cations Comparison (Detection System)

Click here to view


In [Table 7], the control console for Siemens scanners are always of only one monitor used for patients data, acquisitions, viewing, clinical software evaluation as well as filming in addition to the workstation but for GE they using two monitors, one for patient data base and the other for acquisitions, viewing, clinical software evaluation and filming, also all of the scanners are using liquid crystal display monitors of 19-21 inches, with image matrix ranging between (512 × 512), (768 × 768) and (1024 × 1024) megapixel. The CT dose index volume and dose length production which express the CT dose are displayed on the screen for all scanners which facilitate in estimating the doses for both staff and patients. Main computer model is distributed between (hp) for GE scanners and Fujitsu for Siemens and the computer operating system distributed between windows XP for Siemens scanners and Linux for GE scanners, the type of CPUs for the computer system are variable there is intel Pentium xeon, and dual core to duo, also the range of CT numbers display for most of scanners is between −1024 and +3071 (HU) with exception of Siemens Somatom definition 64 slices which has CT number range from −10,240 to +30,710, image storage show that the maximum hard disk capacity is ranged between 40 (G bytes) and 30 × 1024 (tetra) for Siemens Somatom emotion 6 though the later scanner is exist in small hospital and there is no rush of patients in the hospital, the hard disk capacity for image storage (in Gbytes), number of images is ranged from 20,000 for GE Hispeed 4 slices to 520,000 Images for Siemens somatom definition as 64 slices.
Table 7 Technical Specifi cations Comparison (Control Console)

Click here to view


Workstations as in [Table 8] are available with all CT scanners of the region except the portable CT cereTom and GE Hispeed NXi pro. Furthermore most of the workstations of the scanners are with two monitor.
Table 8 Technical Specifi cations Comparison (Independent Work Station)

Click here to view


[Table 9] declares that all CT units of the region have an advanced reconstruction and 3D display with exception to GE HiSpeed, which has only multi-planar reconstruction and Siemens emotion 1 slice in which there is no advanced reconstruction and 3D display as its single slice.
Table 9 Technical Specifi cations Comparison (Advanced Reconstruction Display)

Click here to view


Clinical application software for advanced scanning [Table 10] are almost available among the region scanners, but most of the applications are not in use this is mainly due to lack of specialty and the poor knowledge of the physician about the efficacy of CT scanners and their poor knowledge about the advanced scanning. For example bone mineral densitometry this software is available in 7 out of the 11 scanners, but the percentage of application is 0%, virtual colonscopy in 8 out of total scanners and also not in use, cardiac imaging is found in 2 out of all but not applicable as well as CT perfusion software, in the region though its available among 8 scanners, dental and CT fluoroscopy as software both of them are available in 8 out of the 11 scanners and they are not applicable, for radiotherapy CT simulation software is neither available nor applicable as there is no radiation therapy center in the region, CT angiography is available in 8 scanners out of the total, with application of 2 out of the total scanners (18%) though it can be available in some scanners from the 3D maximum intensity projection.
Table 10: Technical Specifi cations Comparison (Clinical Application Software)

Click here to view


CT scanner markets are driven by the trend towards multi-slice scanners; the global market for CT scanners was valued at $3.7 billion in 2012. Total market value is expected to reach $6 billion by 2019. [4]

CT scanners are purchased in general either to supplement an older scanner to meet the increased demands on services or to take advantage of new developments that enable improved diagnostic faster throughput or any other clinical benefits, CT scanners are, however expensive both to purchase and to operate, hence it's important to select the scanner which offers greatest value of money.

Purchasers are to select scanners that meet their needs of the clinical environment(s) for which it is intended. The significant challenge for buyers is the fact that - even though manufacturers follow international standards when specifying CT performance - different companies' specifications are not necessarily stated in the same terms. Therefore, buyers often need to obtain further information and conduct additional analysis.


  Conclusion Top


Jazan region scanners have a high capability in clinical software with less application. Technical specification of CT scanner was found to be within the standard criteria and it is one of the most important aspects to be considered by buyers in addition to operation and economic considerations.

 
  References Top

1.IAEA Human Health Series No. 19 ,Vienna, 2012: Quality assurance program for computed tomography: diagnostic and therapy application. Publishers are [IAEA] available at: http://www.iaea.org/Publications/index.html.  Back to cited text no. 1
    
2.ICRP Publication 87. Managing Patient Dose in Computed Tomography. Volume 30 NO (4) 2000. Task Group: M.M. Rehani, G. Bongartz, S.J. Golding, L. Gordon, W. Kalender, T. Murakami, P. Shrimpton,R. Albrecht, K. Wei. available at: http://www.icrp.org/publication.asp?id=ICRP%20Publication%2087  Back to cited text no. 2
    
3.NCRP report #147. Structural Shielding Design for Medical X-Ray Imaging Facilities, by Benjamin R. Archer and Joel E. Gray and others November 19, 2004 available at: http://www.ncrppublications.org/Reports/147.  Back to cited text no. 3
    
4.Computed Tomography (CT): Market Shares, Strategies, and Forecasts, Worldwide, 2013 to 2018. by Ellen T. Curtiss and Susan Eustis, WinterGreen research INC report no SH25453932. Lexington, Massachusetts (February 7, 2013) available at: www.wintergreenresearch.com/reports/CT htm. Authors name: Ellen T. Curtiss and Susan Eustis Publishers: WinterGreen research INC report no SH25453932.  Back to cited text no. 4
    
5.Technical Specifications for CT Scanners | eHow.com. By Anastasia Zoldak, available at: http://www.ehow.com/list_7384251_technical-specifications-ct-scanners.html.  Back to cited text no. 5
    
6.Radiology Departments in Public Hospitals of Jazan region and the General Directorate of Jazan Health Affairs.  Back to cited text no. 6
    
7.ImPACT CT scanner evaluation group. CentreforEvidence-based Purchchasing (CEP). Comparative Specifications-16slice CT scanners, guide CEP08025; NHS PASA. March 2009 available at: http://www.impactscan.org/reports/CEP08025.htm.  Back to cited text no. 7
    
8.ImPACT CT scanner evaluation group. CentreforEvidence-based Purchchasing (CEP). 32 to 64 slice CT scanner comparison report version 13 No.05068.september-2005 available at: http://www.impactscan.org/reports/Report05068.htm.  Back to cited text no. 8
    
9.ImPACT CT scanner evaluation group. CentreforEvidence-based Purchchasing (CEP). Multislice CT scanners. Buyers, guide CEP08007; NHS PASA March 2009 available at: http://www.impactscan.org/reports/CEP08007.htm.  Back to cited text no. 9
    
10.Old and outdated medical equipment by Nadeem Esmail, the rules from the European Coordination Committee of the Radiological and Electromedical Industries (ECCREI), June 2011 available at: http://www.fraserinstitute.org/uploadedFiles/fraser-ca/Content/research-news/research/articles/old-and-outdated-medical-equipment.pdf  Back to cited text no. 10
    
11.[UNSCEAR] 2008 report page (60) noted in annex (A) sources and effects of ionizing radiation-volume (1) UN, NY 2010: sources from the report of [OECD] health data 2007 (Organization for Economic Co-operation and Development) figure (B -11): the worldwide last updated number of CT scanners in relation to population (OECD) countries: 28 Oct 2011). available at: http://www.unscear.org/docs/reports/2008/0986753_Report_2008_Annex_A.pdf.  Back to cited text no. 11
    



 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4], [Table 5], [Table 6], [Table 7], [Table 8], [Table 9], [Table 10]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Materials and Me...
Results
Discussion
Conclusion
References
Article Tables

 Article Access Statistics
    Viewed2806    
    Printed42    
    Emailed0    
    PDF Downloaded213    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]